1)Insulin test kits
2)Abbr. GA test kits
3)Enzymatic method
2014 up-to-date high-pricision good quality
glycosylated albumin test kits
Methodology: Enzymatic method
packing: 220 ML R1:(2*40ML) R2:(2*10ML) R3:(3*40ML) Calibrator:2*1ML
We can provide Hitachi,Beckman,Roche special package,bulk packing is also available
Design: OEM welcome
Within Run Precision(S r ) (μmol/L)
control level1:198 μmol/L GA | Control Level2:750 μmol/L GA | Serum Level 1: 241 μmol/L GA | Serum Level 2: 375 μmol/L GA | |
Number of data points | 80 | 80 | 80 | 80 |
Mean | 204 | 751 | 251 | 373 |
SD | 2.15 | 4.91 | 1.94 | 2.41 |
CV(%) | 1.1 | 0.7 | 0.8 | 0.6 |
Between Group Precision (μmol/L)
control level1:198 μmol/L GA | Control Level2:750 μmol/L GA | Serum Level 1: 241 μmol/L GA | Serum Level 2: 375 μmol/L GA | |
Number of data points | 80 | 80 | 80 | 80 |
Mean | 204 | 751 | 251 | 373 |
SD | 2.39 | 5.59 | 3.18 | 3.75 |
CV(%) | 1.2 | 0.7 | 1.3 | 1.0 |
Linearity :21.0~1200.0 μmol/L,r 2 ≥0.990
No interference was observed by TG up to 20mmol/L,Bilirubin 320 μmol/L ,hemoglobin up to 1.6g/L,ascorbic Acid up to 0.2 mmol/L,blood glucose up to 20g/L,Uric Acid up to 36mg/dL
Properties: |
Other
|
Place of Origin: |
Zhejiang China (Mainland)
|
Brand Name: |
MedicalSystem
|
Model Number: |
MS-GA
|
Item name: |
glycosylated albumin test kits
|
Abbr.: |
GA test kits
|
Methodology: |
Enzymatic method
|
Package: |
Hitachi,Beckman,Roche and big package
|
Transportation: |
by Express,by air,by sea
|
Sample delivery: |
about 5 days
|
Cargo delivery: |
15 days after get the payment
|
Sample volume: |
20μl
|
reagent volume: |
R1:200μl,R2:50μl
|